CONFIDENSE™ Module
The CONFIDENSE™ Module is a collection of features that aid in the acquisition and understanding of CARTO™ 3 Maps. The collection of features include Continuous Mapping feature, Tissue Proximity Technology, Wavefront Mapping Annotation, and Map Consistency Display.
Features & benefits

Continuous Mapping
The Continuous Mapping Feature enables continuous acquisition of EA points which meet a set of pre-defined filters.

Tissue Proximity Technology
With Tissue Proximity Technology, the CARTO™ 3 System uses impedance measurements to determine an electrode's proximity to cardiac tissue. If the electrode is in the proximity of the tissue, the electrode is considered to be touching. The proximity status can be used for
the following features:
- Tissue Proximity Indication in the map viewers, Point List, and Selected Point Viewer
- MEM point acquisition based on Tissue Proximity Filter

Wavefront Mapping
The Wavefront Mapping Annotation uses an algorithm to search for the greatest negative deflection (-dV/dt) in each distal unipolar signal and mark it for the local activation time calculation.

Map Consistency Display
The Map Consistency Display is a point filter that identifies points whose Local Activation
Time (LAT) values are not consistent with the LAT values of neighboring points. When this filter is applied, the CARTO™ 3 System marks these outliers.
An integrated workflow

OCTARAY™ Mapping Catheter with TRUEref™ Technology

OPTRELL™ Mapping Catheter

VARIPULSE™ Catheter
Related products

CARTO ELEVATE™ Module
CARTO™ 3 System V8 brings advanced mapping capabilities with enhanced signal analysis tools, improved substrate characterization and utilization of ultrasound technology.
.jpg?width=3840&quality=90&format=webp)
CARTO™ LAM Module

CARTOSOUND™ FAM Module
Powered by Artificial Intelligence, the CARTOSOUND™ FAM Module reconstructs the left atrial (LA) anatomy on the CARTO™ 3 System with a few rotations of your ultrasound catheter, eliminating the need for manual contouring.